Hamamatsu source

From Nucwiki
Revision as of 23:10, 1 June 2022 by Mcmaster (talk | contribs) (Pony DR2 controller)
Jump to: navigation, search

Head controllers are yellow

Hamamatsu Photonics Softex L7452 X-Ray Tube Unit L7452

HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)

Pony DR2 controller

"for PXG-090W" Testing with L7902 tube (100 kV, 250 uA, 24V, 100W)

https://www.ebay.com/itm/152627561420

ROM implies this is a rebranded Hamamatsu C8033-11

note other listing: "Hamamatsu L7901 (C8033) X-RAY100 Control Unit"

Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on


Bring up summary (Pony DR2):

  • Unit off
  • Close both interlocks
  • Rotary switch left / "OFF"
  • Turn key to power on
  • Interlock light should be yellow indicating both interlocks are closed
  • Move rotary switch to middle / "ON"
  • Will slowly warm up to 20 kV. Voltage will be displayed
  • Will sit here for 5 to 10 minutes. AGING light should be on
  • Will slow start ramping up voltage past 20 kV. I started detecting x-rays at this point, although I shut off only at maybe 35 kV


serial

25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter


interlock1

2 wires. Came with shorting plug

Internally appear to go to J1

Why isn't interlock1 and interlock2 in series? Do they serve different purposes?

interlock2

4 wires used in case

  • 1: red
  • 2: orange
  • 3: white
  • 4: gray

Red and orange go to J11 near large MATSUSHITA relay

Think white and gray go to J9 along with a few other wires

Came with adapter cable HCN21A

  • black
  • red
  • white
  • white
  • male pins
  • goes to another cable
  • 12V across white-white
  • shorted white-white and interlock light illuminated!


Internals

SW1: all LEDs on front light when pressed. No serial output

MCU: TMP96C141BF

   $ strings c8033-11_v1.0.5.p9_2012-01-06.bin 
   FOXf
   ??n1
   Micro Foucus X-Ray 130kV Controller
   Version 1.03p 00/03/25 sat.
   KOU PLANNING  CO.LTD.
   @h1!
   ?9sp
   ?6w^
   ?9cI
   ?9c8
   ?9c'
   HIV CUR PHV PCU AGN RST
   ERR 
   ERROR
   X-RAY ON
   X-RAY OFF
   HIV 
   CUR 
   PHV 
   PCU 
   OVER
   NORMAL
   INTER LOCK 
   AGING 
   AGING
   READY
   NOT READY
   BUSY
   REMOTE
   LOCAL
   RESPONSE 
   STANDBY
   CAMERA POWER 
   SMALL
   LARGE
   Micro Foucus X-Ray Controller
   Version 1.03p 00/03/25 sat. 
   KOU PLANNING  CO.LTD.
   Control commands
   HIV_0-1300    CUR_0-300
   PHV_0-1300    PCU_0-300
   AGN_N_N       XON    XOF
   CXT    RST    CFL    CFS
   Status commands
   STX    SHV    SCU    SPV
   SPC    SOV    SIN    SAG
   STS    SRB    SRL    DAG
   9;1d
   DRSLVMFHGN!BTO$x
   x=	x
   gH[;(
   ?.fE
   <8Db
   @0000
    address : +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +A +B +C +D +E +F  0123456789ABCDEF
   ----------------------------------------------------------  ----------------
      PCR      XWA      XBC      XDE      XHL      XIX      XIY      XIZ      XSP
   Version 0.04a  94/06/02 thu.
   ERROR
    xwa:
    xbc:
    xde:
    xhl:
    xix:
    xiy:
    xiz:
    xsp:
    pcr:
     sr:
      spend:
   usec
   TLCS-900 Remote Debugger  Version 0.04a
   KOU PLANNING CO. LTD.     94/06/02 thu.
   D[(start),(end)]       :  memory dump
   R                      :  register display
   S[add]                 :  set memory
   G[(add),(break)]       :  jump to add.
   N[(add),(break)]       :  jump to add.
   L[offset add]          :  load file
   M(source),(end),(dist) :  move memory
   F(start),(end),(data)  :  fill memory
   H                      :  help massage
   V                      :  version
   ?.fE