Faxitron
Faxitron Bioptics
Model | Option | x-ray
Spot size |
x-ray
kVp |
x-ray
mA |
Panel
size |
Panel
resolution (lp/mm) |
Panel
interface |
Mass (kg) | Note |
---|---|---|---|---|---|---|---|---|---|
43855A
43855B 43856A |
0.05-0.5 | 10-130 | 0.3-3 | 200-726 | 64/125/91 cm max focal distance
Be x-ray window
Standard: 10 - 110 kVp, 3 mA, 0.5 mm spot A04: 10 - 130 kVp, 3 mA, 0.5 mm spot M55: 10 - 55 kVp, 0.3 mA, 0.05 mm spot M110: 10 - 110 kVp, 0.3 mA, 0.07 mm spot
21 x 26 39 x 45 30 x 36 | ||||
DX-40 | |||||||||
DX-50 | 0.05 | 10-35 | 0.1 | 5x5 cm
5x10 cm |
10 | USB? | 32? | ||
MX-20 | 0.02 | 10-35 | 0.3 | 5x5 cm
12x12 cm |
normal: 50
contact: 10 |
unit: 68.2
w/ cart: 163.7 |
|||
VersaVision | 21 |
Misc info from: http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html
http://www.rtftechnologies.org/physics/faxitron-DX50-teardown.html
Python code: https://github.com/JohnDMcMaster/faxitron
Faxitron X-Ray DX-50 Digital Biopsy System with Option DC-4 Specimen Radiography
The Faxitron MX-20 Digital Specimen Radiography System is substantially
equivalent to the following to the following predicate devices:
Faxitron Micro 50 Specimen Radiography System-K852634
Faxitron DX-50 Digital Radiography System-K061361
MX-20 510(k) Submission
Page 9 of 30K079
piXarray Digital Specimen Radiography System-K052433
Contents
Cabinet X-ray Systems
https://epic.awi.de/id/eprint/44382/3/Faxitron-tech-manual.pdf
Models 43855A, 43855B and 43856A
Up to 110kVp depending on option
50-500 um spot size
"designed to give high resolution radiographs of small to medium-size objects of all types"
DX-40
DX-50
http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html
- Very detailed and valuable info
- See for manuals and other info
Weight
- DX-50 sample shipping weight w/ box: 67 lb
- DX-50 w/ computer: about 81 lb
Circa 2008
https://www.youtube.com/watch?v=XGCvC7F24ws
- Hammimatsu flat panel detector
- Carbon fiber x-ray window to sensor
https://www.youtube.com/watch?v=2-vQt9GNaIk
Manual option info
- DC2: Dalsa, CCD type. 50 x 50 mm (2" x 2") field of view.
- DC4: Dalsa, CCD type. 50 x 100 mm (2" x 4") field of view.
- DC22: Bioptics, CMOS type. 50 x 50 mm (2" x 2") field of view.
- DC24: Bioptics, CMOS type. 50 x 100 mm (2" x 4") field of view.
- DC44: Bioptics, CMOS type. 100 x 100 mm (4" x 4") field of view.
- DC3: Hamamatsu, CCD type. 24 x 34 mm (0.94" x 1.3") field of view.
- DC5: Hamamatsu, CMOS type. 50 x 50 mm (2" x 2") field of view.
- DC10: Hamamatsu, CMOS type. 100 x 100 mm (4" x 4") field of view.
- DC12: Hamamatsu, CMOS type. 120 x 120 mm (4.7" x 4.7") field of view.
MX-20
Cabinet x-ray
Camera
- DC-12: 12 cm x 12 cm
- DC-5: 5 cm x 5 cm, less expensive
10 lp/mm spatial resolution in contact mode. Magnification allows for spatial resolution of up to 50 lp/mm. The 20 μm nominal focal spot and up to 5X additional geometric magnification
Specs
- Window: 0.01” beryllium
- Resolution: Up to 50 lp/mm (10 lp/mm contact mode)
- 10-35 kVp @ 300 μA
- <20 μm spot size
- Image acquisition time: 15 sec
https://www.eevblog.com/forum/reviews/faxitron-mx-20-x-ray-system-teardown/
https://www.eevblog.com/forum/reviews/faxitron-mx-20-reverse-engineering-(with-some-good-results)/
https://www.youtube.com/watch?v=mYrjOIiJv88
VersaVision
Circa 2017?
Perkin-elmer_xrd-1640_an19-ets-lp
"The MX-20 had an Perkin Elmer panel sensor with parallel output to an SCSI PC interface board. It can do medium sized boards, up to 10" x 12"...The P-E sensor that came with my MX-20 was not it's original sensor. My MX-20 was non-digital, just a film based one (or phosphor plate based). The P-E had a low voltage differential, parallel, output that was sent through a cable to a PCI frame capture board. The board (made by Perkin Elmer) looked like a SCSI interface to the PC. The sensors are available on eBay, sometimes for a good price, but I've not seen a complete system, sensor, power supply, cables, interface card, and software, in a long time. A sensor is on eBay now, see"
tvrsales2 https://www.ebay.com/itm/PERKIN-ELMER-DIGITAL-X-RAY-DETECTOR-SYSTEM-XRD-1640/233202476578
PERKIN ELMER DIGITAL X-RAY DETECTOR SYSTEM XRD-1640
Teledyne dalsa CD26M112 X1
Dalsa Teledyne Xineos CMOS Xray Detector 130mm X 130mm CD26M112 X1
Serial
Faxitron Serial Commands for MX-20 and DX-50 all commands <!xxxxx> must be followed by a <CR>, 'C' and 'A' do not require a <CR> The DX-50 will occasionally miss commands and queries, continue sending the string until the unit responds -------------------------
9600 bps 8 data bits 1 stop bit No parity No handshake
!MR Set mode remote !MR Set mode front panel !V26 Set kV (10-35) !T0140 Set 14.0 sec exposure
Range 0.1-999.9 (T0001-T9999)
!B Initiates X-ray
Then this sequence: Machines replies "X" X-ray Write "C" Continue to start actual X-ray Machines replies "P" Processing Write "A" to abort Machine replies "S" when complete
?S Get state.
Reply ?SW: Warming up Reply ?SD: Door open Reply ?SR: Ready
?M Get mode.
Reply ?MF: Front panel mode Reply ?MR: Remote mode
?D Get device.
Reply MX-20
?R Get revision.
Reply 4.2
?V Get kV.
Reply ?V26
?T Get exposure time.
Reply T0140
Misc
Jul 31, 2018 Hologic acquired Faxitron Bioptics for $85,000,000