Difference between revisions of "Hitachi Vortex"
(Tag: Visual edit) |
(Tag: Visual edit) |
||
Line 436: | Line 436: | ||
|} | |} | ||
What if I re-measure vs 6/10? | What if I re-measure vs 6/10? | ||
+ | |||
<br /> | <br /> | ||
+ | |||
+ | == 2022-09-18 == | ||
+ | |||
+ | |||
+ | Pots are just voltage dividers: | ||
+ | |||
+ | * CW => less negative (decrease magnitude / increase voltage) | ||
+ | * CCW => more negative (increase magnitude / decrease voltage) | ||
+ | * 1 => 100x | ||
+ | * A => 7 | ||
+ | * B => 8 | ||
+ | * C => 14 | ||
+ | * D => 15 | ||
+ | |||
+ | |||
+ | Pot D is responding...wasn't it not last time? Looks unstable though. Might need some repair |
Revision as of 01:01, 19 September 2022
Bargain SDD module McMaster bought on eBay. Can I get it working without too much effort?
Can get a complete unit + PS for $1500-2500
Vendor page: https://www.hitachi-hightech.com/hhs-us/product_list/?ld=sms1&md=sms1-19
Contents
McMaster Vortex-EX
HITATCHI Vortex-EX
Based on vendor images looks like Vortex-60EX
Label:
VORTEX-EX S/N HHSUS-T1344S-302387NF
HITATCHI A 0 B -0.80V C -1.30V D -2.20V SDD Thickness 1000 um
840-266000 Rev AA SER. NO. 0616301
Thermo SCIENTIFIC 5225 Verona Rd Madison, WI 53711
eBay usa.88
https://www.ebay.com/itm/174805037510
Comes w/ power supply and cable
Power supply:
VORTEX ELECTRONICS SII STATUS IO RATE DETECTOR TEC
Detector
SII Vortex A 0 B -0.60V C -1.40V D -1.60V SDD Thickness 500 um Warning: Detector unit will work only with the DPP/PS (serial #) assigned as a System T=1.2V POWER
maximusenergy
https://www.ebay.com/itm/143861496023
Hitachi Vortex-90EX
Has paperwork with it and looks generally complete
New unit
Kit is missing the collimator. Contains all other parts.
085-051000 CBL Vortx ac input
15 pin computer cable
085-050900 CBL Vortx SDD dector
A04357 software license
Hitachi Vortex EX Berylieum
Power supply
Thermo power supplies
Some of these on eBay. Are they related?
DA-15, but looks to be the standard preamplifier found on other LN2 detectors. Is this all these detectors need? What about the A/B/C/D tuning seen on the maximusenergy power supply?
Output Power
5V DC 1.3A
5V DC 1.3A
15V DC 100mA
-15V DC 100mA
-200V DC 6mA
2022-10-04
Inventory:
- VORTEX ELECTRONICS SII controller + matching detector. Detector might have TEC issue / doesn't go down to temperature?
- Spare detector (broken window, disassembled)
- SEM detector in the mail
Open spare detector up and take some pictures. Wire bonding / complex structure
Mystery component is probably ion pump. Its only used at the factory and has external magnet + electronics to drive it not included in the final unit
2022-10-05
Goal: use breakout board to document DA15 pinout a bit
Use software to test first: Prospect 0.1.16 w/ Saturn processor. Tweak settings and can get noise but no real signal
Controller: SII NanoTechnology USA Inc, VORTEX ELECTRONICS
Detector this was paired with:
- A 0V
- B -0.60V
- But test point actually read -0.476 => corrected
- C -1.40V
- D -1.60V
- Turning pot doesn't affect cal voltage
- But voltage is correct so leave alone?
- SDD Thickness 1000 um
- Affects HV?
Pin | No detector
vs pin 1 (V) |
Notes |
---|---|---|
1 | 0.0 | Not same net as 2 |
2 | 0.0 | |
3 | -11.92 | 4, 5, 6, 10 respond with short beep |
4 | +11.90 | |
5 | +11.37 | |
6 | Ground (chassis) | |
7 | Pulse? AC? | |
8 | -46.6 | |
9 | Pulse? AC? | |
10 | Ground (chassis) | |
11 | Pulse? AC? | |
12 | Pulse? AC? | |
13 | Pulse? AC? | |
14 | -137.9 | |
15 | -160.1 |
What if I re-measure vs 6/10?
Pin | Function | Wire color | To harness | Notes |
---|---|---|---|---|
1 | N/A | N/C | ||
2 | N/A | N/C | ||
3 | -12V | Black | ||
4 | +12V | Red | ||
5 | Brown | |||
6 | Ground (chassis) | Green | ||
7 | N/A | N/C | ||
8 | Orange | R01262RevB.XP1 | ||
9 | Ground (chassis) | Green | ||
10 | Ground (chassis) | Green | ||
11 | Ground (chassis) | Green | ||
12 | N/A | N/C | ||
13 | N/A | N/C | ||
14 | Yellow | R01262RevB.XP1 | ||
15 | Gray | R01262RevB.XP1 |
Pin | Function | Notes |
---|---|---|
1 | ||
2 | ||
3 | ||
4 | ||
5 | ||
6 | ||
7 | ||
8 | ||
9 | ||
10 | ||
11 | ||
12 | ||
13 | ||
14 | ||
15 |
Pin | Function | Notes |
---|---|---|
1 | N/C | |
2 | N/C | |
3 | -12V | |
4 | +12V | |
5 | ? | FIXME
Reference: +11.37V |
6 | Ground | |
7 | N/C | |
8 | HV1 | Reference: -46.6V |
9 | Ground | |
10 | Ground | |
11 | Ground | |
12 | N/C | |
13 | N/C | |
14 | HV2 | Reference: -137.9V |
15 | HV3 | Reference: -160.1V |
What if I re-measure vs 6/10?
2022-09-18
Pots are just voltage dividers:
- CW => less negative (decrease magnitude / increase voltage)
- CCW => more negative (increase magnitude / decrease voltage)
- 1 => 100x
- A => 7
- B => 8
- C => 14
- D => 15
Pot D is responding...wasn't it not last time? Looks unstable though. Might need some repair