Difference between revisions of "Hamamatsu source"
Line 5: | Line 5: | ||
HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller) | HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller) | ||
− | + | = Pony controller = | |
"Pony Hamamatsu DR2 X-ray System X-RAYCONTROL Unit for PXG-090W & Cables" | "Pony Hamamatsu DR2 X-ray System X-RAYCONTROL Unit for PXG-090W & Cables" | ||
Line 12: | Line 12: | ||
− | + | == serial == | |
25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter | 25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter | ||
− | + | == interlock1 == | |
2 wires. Came with shorting plug | 2 wires. Came with shorting plug | ||
Line 25: | Line 25: | ||
Why isn't interlock1 and interlock2 in series? Do they serve different purposes? | Why isn't interlock1 and interlock2 in series? Do they serve different purposes? | ||
− | + | == interlock2 == | |
4 wires used in case | 4 wires used in case | ||
Line 38: | Line 38: | ||
− | + | == Internals == | |
SW1: all LEDs on front light when pressed. No serial output | SW1: all LEDs on front light when pressed. No serial output | ||
− | |||
− | |||
MCU: TMP96C141BF | MCU: TMP96C141BF |
Revision as of 21:56, 1 June 2022
Head controllers are yellow
Hamamatsu Photonics Softex L7452 X-Ray Tube Unit L7452
HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)
Pony controller
"Pony Hamamatsu DR2 X-ray System X-RAYCONTROL Unit for PXG-090W & Cables"
Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on
serial
25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter
interlock1
2 wires. Came with shorting plug
Internally appear to go to J1
Why isn't interlock1 and interlock2 in series? Do they serve different purposes?
interlock2
4 wires used in case
- 1: red
- 2: orange
- 3: white
- 4: gray
Red and orange go to J11 near large MATSUSHITA relay
Think white and gray go to J9 along with a few other wires
Internals
SW1: all LEDs on front light when pressed. No serial output
MCU: TMP96C141BF
$ strings c8033-11_v1.0.5.p9_2012-01-06.bin FOXf ??n1 Micro Foucus X-Ray 130kV Controller Version 1.03p 00/03/25 sat. KOU PLANNING CO.LTD. @h1! ?9sp ?6w^ ?9cI ?9c8 ?9c' HIV CUR PHV PCU AGN RST ERR ERROR X-RAY ON X-RAY OFF HIV CUR PHV PCU OVER NORMAL INTER LOCK AGING AGING READY NOT READY BUSY REMOTE LOCAL RESPONSE STANDBY CAMERA POWER SMALL LARGE Micro Foucus X-Ray Controller Version 1.03p 00/03/25 sat. KOU PLANNING CO.LTD. Control commands HIV_0-1300 CUR_0-300 PHV_0-1300 PCU_0-300 AGN_N_N XON XOF CXT RST CFL CFS Status commands STX SHV SCU SPV SPC SOV SIN SAG STS SRB SRL DAG 9;1d DRSLVMFHGN!BTO$x x= x gH[;( ?.fE <8Db @0000 address : +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +A +B +C +D +E +F 0123456789ABCDEF ---------------------------------------------------------- ---------------- PCR XWA XBC XDE XHL XIX XIY XIZ XSP Version 0.04a 94/06/02 thu. ERROR xwa: xbc: xde: xhl: xix: xiy: xiz: xsp: pcr: sr: spend: usec TLCS-900 Remote Debugger Version 0.04a KOU PLANNING CO. LTD. 94/06/02 thu. D[(start),(end)] : memory dump R : register display S[add] : set memory G[(add),(break)] : jump to add. N[(add),(break)] : jump to add. L[offset add] : load file M(source),(end),(dist) : move memory F(start),(end),(data) : fill memory H : help massage V : version ?.fE