Difference between revisions of "Faxitron"

From Nucwiki
Jump to: navigation, search
(Serial)
Line 115: Line 115:
  
  
Misc info from: <nowiki>http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html</nowiki>
+
Python code: https://github.com/JohnDMcMaster/faxitron
  
<nowiki>http://www.rtftechnologies.org/physics/faxitron-DX50-teardown.html</nowiki>
+
Misc info from:
  
Python code: <nowiki>https://github.com/JohnDMcMaster/faxitron</nowiki>
+
* http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html
 +
* http://www.rtftechnologies.org/physics/faxitron-DX50-teardown.html
  
  

Revision as of 16:24, 4 January 2020

Faxitron Bioptics

Model Family Option x-ray

Spot size (um)

x-ray

kVp

x-ray

mA

Panel

size

Panel

resolution

(lp/mm)

Panel

interface

Mass (kg) Note
43855A

43855B

43856A

50 - 500 10-130 0.3-3 200-726 64/125/91 cm max focal distance

Be x-ray window


Tube options:

Standard: 10 - 110 kVp, 3 mA, 0.5 mm spot

A04: 10 - 130 kVp, 3 mA, 0.5 mm spot

M55: 10 - 55 kVp, 0.3 mA, 0.05 mm spot

M110: 10 - 110 kVp, 0.3 mA, 0.07 mm spot


Panel size? maybe

21 x 26

39 x 45

30 x 36

DX-50 DX-50 50 10-35 0.1 5x5 cm

5x10 cm

10 USB? 32?
MX-20 DX-50 20 10-35 0.3 5x5 cm

12x12 cm

normal: 50

contact: 10

unit: 68.2

w/ cart: 163.7

LX-60 DX-50 20 max

(10-12 typ)

10-60 0.3 12x12 cm 10 150
VersaVision 21


Python code: https://github.com/JohnDMcMaster/faxitron

Misc info from:


Faxitron X-Ray DX-50 Digital Biopsy System with Option DC-4 Specimen Radiography

The Faxitron MX-20 Digital Specimen Radiography System is substantially

equivalent to the following to the following predicate devices:

Faxitron Micro 50 Specimen Radiography System-K852634

Faxitron DX-50 Digital Radiography System-K061361

MX-20 510(k) Submission

Page 9 of 30K079

piXarray Digital Specimen Radiography System-K052433


Cabinet X-ray Systems

https://epic.awi.de/id/eprint/44382/3/Faxitron-tech-manual.pdf

Models 43855A, 43855B and 43856A

Up to 110kVp depending on option

50-500 um spot size

"designed to give high resolution radiographs of small to medium-size objects of all types"

DX-40

DX-50

http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html

  • Very detailed and valuable info
  • See for manuals and other info

Weight

  • DX-50 sample shipping weight w/ box: 67 lb
  • DX-50 w/ computer: about 81 lb


Circa 2008

https://www.youtube.com/watch?v=XGCvC7F24ws

  • Hammimatsu flat panel detector
  • Carbon fiber x-ray window to sensor

https://www.youtube.com/watch?v=2-vQt9GNaIk


Manual option info

  • DC2: Dalsa, CCD type. 50 x 50 mm (2" x 2") field of view.
  • DC4: Dalsa, CCD type. 50 x 100 mm (2" x 4") field of view.
  • DC22: Bioptics, CMOS type. 50 x 50 mm (2" x 2") field of view.
  • DC24: Bioptics, CMOS type. 50 x 100 mm (2" x 4") field of view.
  • DC44: Bioptics, CMOS type. 100 x 100 mm (4" x 4") field of view.
  • DC3: Hamamatsu, CCD type. 24 x 34 mm (0.94" x 1.3") field of view.
  • DC5: Hamamatsu, CMOS type. 50 x 50 mm (2" x 2") field of view.
  • DC10: Hamamatsu, CMOS type. 100 x 100 mm (4" x 4") field of view.
  • DC12: Hamamatsu, CMOS type. 120 x 120 mm (4.7" x 4.7") field of view.

MX-20

Cabinet x-ray

Spec sheet

Camera

  • DC-12: 12 cm x 12 cm
  • DC-5: 5 cm x 5 cm, less expensive

10 lp/mm spatial resolution in contact mode. Magnification allows for spatial resolution of up to 50 lp/mm. The 20 μm nominal focal spot and up to 5X additional geometric magnification

Specs

  • Window: 0.01” beryllium
  • Resolution: Up to 50 lp/mm (10 lp/mm contact mode)
  • 10-35 kVp @ 300 μA
  • <20 μm spot size
  • Image acquisition time: 15 sec


https://www.eevblog.com/forum/reviews/faxitron-mx-20-x-ray-system-teardown/

https://www.eevblog.com/forum/reviews/faxitron-mx-20-reverse-engineering-(with-some-good-results)/


https://www.youtube.com/watch?v=mYrjOIiJv88

VersaVision

Circa 2017?

Perkin-elmer_xrd-1640_an19-ets-lp

"The MX-20 had an Perkin Elmer panel sensor with parallel output to an SCSI PC interface board. It can do medium sized boards, up to 10" x 12"...The P-E sensor that came with my MX-20 was not it's original sensor. My MX-20 was non-digital, just a film based one (or phosphor plate based). The P-E had a low voltage differential, parallel, output that was sent through a cable to a PCI frame capture board. The board (made by Perkin Elmer) looked like a SCSI interface to the PC. The sensors are available on eBay, sometimes for a good price, but I've not seen a complete system, sensor, power supply, cables, interface card, and software, in a long time. A sensor is on eBay now, see"

tvrsales2 https://www.ebay.com/itm/PERKIN-ELMER-DIGITAL-X-RAY-DETECTOR-SYSTEM-XRD-1640/233202476578

PERKIN ELMER DIGITAL X-RAY DETECTOR SYSTEM XRD-1640

Perkin-elmer xrd-1640 an19-ets-lp 0.jpg Perkin-elmer xrd-1640 an19-ets-lp 1.jpg Perkin-elmer xrd-1640 an19-ets-lp 2.jpg Perkin-elmer xrd-1640 an19-ets-lp 4.jpg Perkin-elmer xrd-1640 an19-ets-lp 6.jpg


Teledyne dalsa CD26M112 X1

Dalsa Teledyne Xineos CMOS Xray Detector 130mm X 130mm CD26M112 X1

Teledyne dalsa cd26m112-x1 srlsales 0.jpg Teledyne dalsa cd26m112-x1 srlsales 1.jpg Teledyne dalsa cd26m112-x1 srlsales 2.jpg

Misc

Jul 31, 2018 Hologic acquired Faxitron Bioptics for $85,000,000