== Serial ==
 
== Serial ==
  
 +
    Faxitron Serial Commands for MX-20 and DX-50
 +
    all commands <!xxxxx> must be followed by a <CR>, 'C' and 'A' do not require a <CR>
 +
    The DX-50 will occasionally miss commands and queries, continue sending the string until the unit responds
 +
    -------------------------
  
 +
    9600 bps
 +
    8 data bits
 +
    1 stop bit
 +
    No parity
 +
    No handshake
 +
 +
    !MR Set mode remote
 +
    !MR Set mode front panel
 +
    !V26 Set kV (10-35)
 +
    !T0140 Set 14.0 sec exposure
 +
    Range 0.1-999.9 (T0001-T9999)
 +
 +
    !B Initiates X-ray
 +
    Then this sequence:
 +
    Machines replies "X" X-ray
 +
    Write "C" Continue to start actual X-ray
 +
    Machines replies "P" Processing
 +
    Write "A" to abort
 +
    Machine replies "S" when complete
 +
 +
    ?S Get state.
 +
    Reply ?SW: Warming up
 +
    Reply ?SD: Door open
 +
    Reply ?SR: Ready
 +
 +
    ?M Get mode.
 +
    Reply ?MF: Front panel mode
 +
    Reply ?MR: Remote mode
 +
 +
    ?D Get device.
 +
    Reply MX-20
 +
 +
    ?R Get revision.
 +
    Reply 4.2
 +
 +
    ?V Get kV.
 +
    Reply ?V26
 +
 +
    ?T Get exposure time.
 +
    Reply T0140
  
<br />
 
 
==Misc==
 
==Misc==
  
 
Jul 31, 2018
 
Jul 31, 2018
 
Hologic acquired Faxitron Bioptics for $85,000,000
 
Hologic acquired Faxitron Bioptics for $85,000,000

Revision as of 23:06, 23 November 2019

Faxitron Bioptics

Model Option x-ray

Spot size

x-ray

kVp

x-ray

mA

Panel

size

Panel

resolution

(lp/mm)

Panel

interface

Mass (kg) Note
43855A

43855B

43856A

0.05-0.5 10-130 0.3-3 200-726 64/125/91 cm max focal distance

Be x-ray window


Tube options:

Standard: 10 - 110 kVp, 3 mA, 0.5 mm spot

A04: 10 - 130 kVp, 3 mA, 0.5 mm spot

M55: 10 - 55 kVp, 0.3 mA, 0.05 mm spot

M110: 10 - 110 kVp, 0.3 mA, 0.07 mm spot


Panel size? maybe

21 x 26

39 x 45

30 x 36

DX-40
DX-50 0.05 10-35 0.1 5x5 cm

5x10 cm

10 USB?
MX-20 0.02 10-35 0.3 5x5 cm

12x12 cm

normal: 50

contact: 10

unit: 68.2

w/ cart: 163.7

VersaVision 21


Misc info from: http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html


Faxitron X-Ray DX-50 Digital Biopsy System with Option DC-4 Specimen Radiography


The Faxitron MX-20 Digital Specimen Radiography System is substantially

equivalent to the following to the following predicate devices:

Faxitron Micro 50 Specimen Radiography System-K852634

Faxitron DX-50 Digital Radiography System-K061361

MX-20 510(k) Submission

Page 9 of 30K079

piXarray Digital Specimen Radiography System-K052433


Cabinet X-ray Systems

https://epic.awi.de/id/eprint/44382/3/Faxitron-tech-manual.pdf

Models 43855A, 43855B and 43856A

Up to 110kVp depending on option

50-500 um spot size

"designed to give high resolution radiographs of small to medium-size objects of all types"

DX-40

DX-50

http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html

  • Very detailed and valuable info
  • See for manuals and other info


Circa 2008

https://www.youtube.com/watch?v=XGCvC7F24ws

  • Hammimatsu flat panel detector
  • Carbon fiber x-ray window to sensor

https://www.youtube.com/watch?v=2-vQt9GNaIk


Manual option info

  • DC2: Dalsa, CCD type. 50 x 50 mm (2" x 2") field of view.
  • DC4: Dalsa, CCD type. 50 x 100 mm (2" x 4") field of view.
  • DC22: Bioptics, CMOS type. 50 x 50 mm (2" x 2") field of view.
  • DC24: Bioptics, CMOS type. 50 x 100 mm (2" x 4") field of view.
  • DC44: Bioptics, CMOS type. 100 x 100 mm (4" x 4") field of view.
  • DC3: Hamamatsu, CCD type. 24 x 34 mm (0.94" x 1.3") field of view.
  • DC5: Hamamatsu, CMOS type. 50 x 50 mm (2" x 2") field of view.
  • DC10: Hamamatsu, CMOS type. 100 x 100 mm (4" x 4") field of view.
  • DC12: Hamamatsu, CMOS type. 120 x 120 mm (4.7" x 4.7") field of view.

MX-20

Cabinet x-ray

Spec sheet

Camera

  • DC-12: 12 cm x 12 cm
  • DC-5: 5 cm x 5 cm, less expensive

10 lp/mm spatial resolution in contact mode. Magnification allows for spatial resolution of up to 50 lp/mm. The 20 μm nominal focal spot and up to 5X additional geometric magnification

Specs

  • Window: 0.01” beryllium
  • Resolution: Up to 50 lp/mm (10 lp/mm contact mode)
  • 10-35 kVp @ 300 μA
  • <20 μm spot size
  • Image acquisition time: 15 sec


https://www.eevblog.com/forum/reviews/faxitron-mx-20-x-ray-system-teardown/

https://www.eevblog.com/forum/reviews/faxitron-mx-20-reverse-engineering-(with-some-good-results)/


https://www.youtube.com/watch?v=mYrjOIiJv88

VersaVision

Circa 2017?

Perkin-elmer_xrd-1640_an19-ets-lp

"The MX-20 had an Perkin Elmer panel sensor with parallel output to an SCSI PC interface board. It can do medium sized boards, up to 10" x 12"...The P-E sensor that came with my MX-20 was not it's original sensor. My MX-20 was non-digital, just a film based one (or phosphor plate based). The P-E had a low voltage differential, parallel, output that was sent through a cable to a PCI frame capture board. The board (made by Perkin Elmer) looked like a SCSI interface to the PC. The sensors are available on eBay, sometimes for a good price, but I've not seen a complete system, sensor, power supply, cables, interface card, and software, in a long time. A sensor is on eBay now, see"

tvrsales2 https://www.ebay.com/itm/PERKIN-ELMER-DIGITAL-X-RAY-DETECTOR-SYSTEM-XRD-1640/233202476578

PERKIN ELMER DIGITAL X-RAY DETECTOR SYSTEM XRD-1640

Perkin-elmer xrd-1640 an19-ets-lp 0.jpg Perkin-elmer xrd-1640 an19-ets-lp 1.jpg Perkin-elmer xrd-1640 an19-ets-lp 2.jpg Perkin-elmer xrd-1640 an19-ets-lp 4.jpg Perkin-elmer xrd-1640 an19-ets-lp 6.jpg


Teledyne dalsa CD26M112 X1

Dalsa Teledyne Xineos CMOS Xray Detector 130mm X 130mm CD26M112 X1

Teledyne dalsa cd26m112-x1 srlsales 0.jpg Teledyne dalsa cd26m112-x1 srlsales 1.jpg Teledyne dalsa cd26m112-x1 srlsales 2.jpg

Serial

   Faxitron Serial Commands for MX-20 and DX-50
   all commands <!xxxxx> must be followed by a <CR>, 'C' and 'A' do not require a <CR>
   The DX-50 will occasionally miss commands and queries, continue sending the string until the unit responds
   -------------------------
   9600 bps
   8 data bits
   1 stop bit
   No parity
   No handshake
   !MR	Set mode remote
   !MR	Set mode front panel
   !V26	Set kV (10-35)
   !T0140	Set 14.0 sec exposure

Range 0.1-999.9 (T0001-T9999)

   !B	Initiates X-ray

Then this sequence: Machines replies "X" X-ray Write "C" Continue to start actual X-ray Machines replies "P" Processing Write "A" to abort Machine replies "S" when complete

   ?S	Get state.

Reply ?SW: Warming up Reply ?SD: Door open Reply ?SR: Ready

   ?M	Get mode.

Reply ?MF: Front panel mode Reply ?MR: Remote mode

   ?D	Get device.

Reply MX-20

   ?R	Get revision.

Reply 4.2

   ?V	Get kV.

Reply ?V26

   ?T	Get exposure time.

Reply T0140

Misc

Jul 31, 2018 Hologic acquired Faxitron Bioptics for $85,000,000