Difference between revisions of "Hitachi Vortex"

From Nucwiki
Jump to: navigation, search
(2022-10-05)
Line 116: Line 116:
 
<br />
 
<br />
 
Use software to test first: Prospect 0.1.16 w/ Saturn processor. Tweak settings and can get noise but no real signal
 
Use software to test first: Prospect 0.1.16 w/ Saturn processor. Tweak settings and can get noise but no real signal
 +
 +
Controller: SII NanoTechnology USA Inc, VORTEX ELECTRONICS
  
  
 
Detector this was paired with:
 
Detector this was paired with:
  
* A 0V
+
*A 0V
* B -0.80V
+
*B -0.80V
* C -1.30V
+
*C -1.30V
* D -2.20V
+
*D -2.20V
* SDD Thickness 1000 um
+
*SDD Thickness 1000 um
** Affects HV?
+
**Affects HV?
  
 
{| class="wikitable"
 
{| class="wikitable"

Revision as of 21:54, 5 September 2022

Bargain SDD module McMaster bought on eBay. Can I get it working without too much effort?

Can get a complete unit + PS for $1500-2500

Vendor page: https://www.hitachi-hightech.com/hhs-us/product_list/?ld=sms1&md=sms1-19

McMaster Vortex-EX

HITATCHI Vortex-EX

Based on vendor images looks like Vortex-60EX

Label:

VORTEX-EX
S/N HHSUS-T1344S-302387NF
HITATCHI
A 0
B -0.80V
C -1.30V
D -2.20V
SDD Thickness 1000 um
840-266000 Rev AA
SER. NO. 0616301
Thermo SCIENTIFIC
5225 Verona Rd
Madison, WI 53711

eBay usa.88

https://www.ebay.com/itm/174805037510

Comes w/ power supply and cable

Power supply:

VORTEX ELECTRONICS
SII

STATUS
IO
RATE
DETECTOR
TEC

Detector

SII Vortex

A 0
B -0.60V
C -1.40V
D -1.60V
SDD Thickness 500 um

Warning: Detector unit will work only with the DPP/PS (serial #) assigned as a System
T=1.2V
POWER

maximusenergy

https://www.ebay.com/itm/143861496023

Hitachi Vortex-90EX

Has paperwork with it and looks generally complete

New unit

Kit is missing the collimator. Contains all other parts.

085-051000 CBL Vortx ac input

15 pin computer cable

085-050900 CBL Vortx SDD dector

A04357 software license

Hitachi Vortex EX Berylieum

Power supply

Thermo power supplies

Some of these on eBay. Are they related?

DA-15, but looks to be the standard preamplifier found on other LN2 detectors. Is this all these detectors need? What about the A/B/C/D tuning seen on the maximusenergy power supply?

Output Power

5V DC 1.3A

5V DC 1.3A

15V DC 100mA

-15V DC 100mA

-200V DC 6mA


2022-10-04

Inventory:

  • VORTEX ELECTRONICS SII controller + matching detector. Detector might have TEC issue / doesn't go down to temperature?
  • Spare detector (broken window, disassembled)
  • SEM detector in the mail

Open spare detector up and take some pictures. Wire bonding / complex structure

Mystery component is probably ion pump. Its only used at the factory and has external magnet + electronics to drive it not included in the final unit

2022-10-05

Goal: use breakout board to document DA15 pinout a bit
Use software to test first: Prospect 0.1.16 w/ Saturn processor. Tweak settings and can get noise but no real signal

Controller: SII NanoTechnology USA Inc, VORTEX ELECTRONICS


Detector this was paired with:

  • A 0V
  • B -0.80V
  • C -1.30V
  • D -2.20V
  • SDD Thickness 1000 um
    • Affects HV?
Power supply w/ N/C detector
Pin No detector

vs pin 1 (V)

Notes
1 0.0 Not same net as 2
2 0.0
3 -11.92 4, 5, 6, 10 respond with short beep
4 +11.90
5 +11.37
6 Ground (chassis)
7 Pulse? AC?
8 -46.6
9 Pulse? AC?
10 Ground (chassis)
11 Pulse? AC?
12 Pulse? AC?
13 Pulse? AC?
14 -137.9
15 -160.1

What if I re-measure vs 6/10?

Broken window unit internal notes
Pin Function Wire color To harness Notes
1 N/A N/C
2 N/A N/C
3 -12V Black
4 +12V Red
5 Brown
6 Ground (chassis) Green
7 N/A N/C
8 Orange R01262RevB.XP1
9 Ground (chassis) Green
10 Ground (chassis) Green
11 Ground (chassis) Green
12 N/A N/C
13 N/A N/C
14 Yellow R01262RevB.XP1
15 Gray R01262RevB.XP1


Controller internals
Pin Function Notes
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15


Aggregate functionality
Pin Function Notes
1 N/C
2 N/C
3 -12V
4 +12V
5 ? FIXME

Reference: +11.37V

6 Ground
7 N/C
8 HV1 Reference: -46.6V
9 Ground
10 Ground
11 Ground
12 N/C
13 N/C
14 HV2 Reference: -137.9V
15 HV3 Reference: -160.1V

What if I re-measure vs 6/10?