HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller) | HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller) | ||
+ | |||
+ | Hamamatsu X-Ray Control Unit L9191(C9193) L9191(C9193) 60Days Warranty | ||
+ | |||
= Pony DR2 controller = | = Pony DR2 controller = | ||
+ | |||
+ | "for PXG-090W" Testing with L7902 tube (100 kV, 250 uA, 24V, 100W) | ||
+ | |||
+ | https://www.ebay.com/itm/152627561420 | ||
+ | |||
+ | https://github.com/JohnDMcMaster/Sea33 | ||
ROM implies this is a rebranded Hamamatsu C8033-11 | ROM implies this is a rebranded Hamamatsu C8033-11 | ||
− | " | + | note other listing: "Hamamatsu L7901 (C8033) X-RAY100 Control Unit" |
Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on | Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on | ||
+ | |||
+ | |||
+ | Bring up summary (Pony DR2): | ||
+ | * Unit off | ||
+ | * Close both interlocks | ||
+ | * Rotary switch left / "OFF" | ||
+ | * Turn key to power on | ||
+ | * Interlock light should be yellow indicating both interlocks are closed | ||
+ | * Move rotary switch to middle / "ON" | ||
+ | * Will slowly warm up to 20 kV. Voltage will be displayed | ||
+ | * Will sit here for 5 to 10 minutes. AGING light should be on | ||
+ | * Will slow start ramping up voltage past 20 kV. I started detecting x-rays at this point, although I shut off only at maybe 35 kV | ||
Line 18: | Line 39: | ||
25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter | 25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter | ||
+ | |||
+ | == x-ray cable == | ||
+ | |||
+ | Controller panel mount | ||
+ | * JAE SR0N2A25-24S | ||
+ | * 24 pin | ||
+ | |||
+ | Cable male | ||
+ | * 25-24 | ||
== interlock1 == | == interlock1 == | ||
Line 141: | Line 171: | ||
V : version | V : version | ||
?.fE | ?.fE | ||
+ | |||
+ | = Small head = | ||
+ | |||
+ | Head panel mount connector: | ||
+ | * JAE SRCN2A25-16P | ||
+ | * 16 pins | ||
+ | * https://www.jae.com/en/connectors/series/detail/product/id=68526&application_code=A1030 | ||
+ | |||
+ | looks like goes with C6973-01 / MICROFOCUS X-RAY UNIT MSI-130 | ||
+ | |||
+ | |||
+ | Shimadzu of interest |
Latest revision as of 10:17, 2 June 2022
Head controllers are yellow
Hamamatsu Photonics Softex L7452 X-Ray Tube Unit L7452
HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)
Hamamatsu X-Ray Control Unit L9191(C9193) L9191(C9193) 60Days Warranty
Contents
Pony DR2 controller
"for PXG-090W" Testing with L7902 tube (100 kV, 250 uA, 24V, 100W)
https://www.ebay.com/itm/152627561420
https://github.com/JohnDMcMaster/Sea33
ROM implies this is a rebranded Hamamatsu C8033-11
note other listing: "Hamamatsu L7901 (C8033) X-RAY100 Control Unit"
Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on
Bring up summary (Pony DR2):
- Unit off
- Close both interlocks
- Rotary switch left / "OFF"
- Turn key to power on
- Interlock light should be yellow indicating both interlocks are closed
- Move rotary switch to middle / "ON"
- Will slowly warm up to 20 kV. Voltage will be displayed
- Will sit here for 5 to 10 minutes. AGING light should be on
- Will slow start ramping up voltage past 20 kV. I started detecting x-rays at this point, although I shut off only at maybe 35 kV
serial
25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter
x-ray cable
Controller panel mount
- JAE SR0N2A25-24S
- 24 pin
Cable male
- 25-24
interlock1
2 wires. Came with shorting plug
Internally appear to go to J1
Why isn't interlock1 and interlock2 in series? Do they serve different purposes?
interlock2
4 wires used in case
- 1: red
- 2: orange
- 3: white
- 4: gray
Red and orange go to J11 near large MATSUSHITA relay
Think white and gray go to J9 along with a few other wires
Came with adapter cable HCN21A
- black
- red
- white
- white
- male pins
- goes to another cable
- 12V across white-white
- shorted white-white and interlock light illuminated!
Internals
SW1: all LEDs on front light when pressed. No serial output
MCU: TMP96C141BF
$ strings c8033-11_v1.0.5.p9_2012-01-06.bin FOXf ??n1 Micro Foucus X-Ray 130kV Controller Version 1.03p 00/03/25 sat. KOU PLANNING CO.LTD. @h1! ?9sp ?6w^ ?9cI ?9c8 ?9c' HIV CUR PHV PCU AGN RST ERR ERROR X-RAY ON X-RAY OFF HIV CUR PHV PCU OVER NORMAL INTER LOCK AGING AGING READY NOT READY BUSY REMOTE LOCAL RESPONSE STANDBY CAMERA POWER SMALL LARGE Micro Foucus X-Ray Controller Version 1.03p 00/03/25 sat. KOU PLANNING CO.LTD. Control commands HIV_0-1300 CUR_0-300 PHV_0-1300 PCU_0-300 AGN_N_N XON XOF CXT RST CFL CFS Status commands STX SHV SCU SPV SPC SOV SIN SAG STS SRB SRL DAG 9;1d DRSLVMFHGN!BTO$x x= x gH[;( ?.fE <8Db @0000 address : +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +A +B +C +D +E +F 0123456789ABCDEF ---------------------------------------------------------- ---------------- PCR XWA XBC XDE XHL XIX XIY XIZ XSP Version 0.04a 94/06/02 thu. ERROR xwa: xbc: xde: xhl: xix: xiy: xiz: xsp: pcr: sr: spend: usec TLCS-900 Remote Debugger Version 0.04a KOU PLANNING CO. LTD. 94/06/02 thu. D[(start),(end)] : memory dump R : register display S[add] : set memory G[(add),(break)] : jump to add. N[(add),(break)] : jump to add. L[offset add] : load file M(source),(end),(dist) : move memory F(start),(end),(data) : fill memory H : help massage V : version ?.fE
Small head
Head panel mount connector:
- JAE SRCN2A25-16P
- 16 pins
- https://www.jae.com/en/connectors/series/detail/product/id=68526&application_code=A1030
looks like goes with C6973-01 / MICROFOCUS X-RAY UNIT MSI-130
Shimadzu of interest