Difference between revisions of "Faxitron"
(Tag: Visual edit) |
(Tag: Visual edit) |
||
Line 221: | Line 221: | ||
{| class="wikitable" | {| class="wikitable" | ||
|+ | |+ | ||
− | ! | + | !Weidmüller-4 |
!Wire color | !Wire color | ||
!Volt | !Volt | ||
Line 250: | Line 250: | ||
!DB9 | !DB9 | ||
!Wire | !Wire | ||
− | ! | + | !Weidmüller-8 |
!Volt | !Volt | ||
|- | |- | ||
Line 318: | Line 318: | ||
| | | | ||
|} | |} | ||
− | + | Weidmüller BL series 3.5 mm connectors | |
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
==MX-20== | ==MX-20== | ||
Cabinet x-ray | Cabinet x-ray |
Revision as of 21:21, 7 January 2020
TODO 2020-01-05: have significant info to add
Faxitron Bioptics
Model | Family | Option | x-ray
Spot size (um) |
x-ray
kVp |
x-ray
mA |
Panel
size |
Panel
resolution (lp/mm) |
Panel
interface |
Mass (kg) | Note |
---|---|---|---|---|---|---|---|---|---|---|
43855A
43855B 43856A |
50 - 500 | 10-130 | 0.3-3 | 200-726 | 64/125/91 cm max focal distance
Be x-ray window
Standard: 10 - 110 kVp, 3 mA, 0.5 mm spot A04: 10 - 130 kVp, 3 mA, 0.5 mm spot M55: 10 - 55 kVp, 0.3 mA, 0.05 mm spot M110: 10 - 110 kVp, 0.3 mA, 0.07 mm spot
21 x 26 39 x 45 30 x 36 | |||||
DX-50 | DX-50 | 50 | 10-35 | 0.1 | 5x5 cm
5x10 cm |
10 | USB? | 32? | ||
MX-20 | DX-50 | 20 | 10-35 | 0.3 | 5x5 cm
12x12 cm |
normal: 50
contact: 10 |
unit: 68.2
w/ cart: 163.7 |
|||
LX-60 | DX-50 | 20 max
(10-12 typ) |
10-60 | 0.3 | 12x12 cm | 10 | 150 | |||
VersaVision | 21 | |||||||||
MultiRad |
See also: Hamamatsu panel
Python code: https://github.com/JohnDMcMaster/faxitron
Alternate DCAM python bindings: https://github.com/lumasullo/Tempesta/tree/master/hamamatsu
Misc info from:
- http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html
- http://www.rtftechnologies.org/physics/faxitron-DX50-teardown.html
Faxitron X-Ray DX-50 Digital Biopsy System with Option DC-4 Specimen Radiography
The Faxitron MX-20 Digital Specimen Radiography System is substantially
equivalent to the following to the following predicate devices:
Faxitron Micro 50 Specimen Radiography System-K852634
Faxitron DX-50 Digital Radiography System-K061361
MX-20 510(k) Submission
Page 9 of 30K079
piXarray Digital Specimen Radiography System-K052433
"The USB Hamamatsu 120mm x 120mm camera array cost £25K on top of the basic MX-20 cabinet cost of £25K. Scary prices." https://www.eevblog.com/forum/buysellwanted/(uk)-faxitron-mx-20-120mm-x-1200mm-complete-system-frasers-own-unit/msg2355963/#msg2355963
Contents
Cabinet X-ray Systems
https://epic.awi.de/id/eprint/44382/3/Faxitron-tech-manual.pdf
Models 43855A, 43855B and 43856A
Up to 110kVp depending on option
50-500 um spot size
"designed to give high resolution radiographs of small to medium-size objects of all types"
DX-40
DX-50
http://www.rtftechnologies.org/physics/faxitron-DX50-CT-scan.html
- Very detailed and valuable info
- See for manuals and other info
Weight
- DX-50 sample shipping weight w/ box: 67 lb
- DX-50 w/ computer: about 81 lb
Circa 2008
https://www.youtube.com/watch?v=XGCvC7F24ws
- Hammimatsu flat panel detector
- Carbon fiber x-ray window to sensor
https://www.youtube.com/watch?v=2-vQt9GNaIk
Manual option info
- DC2: Dalsa, CCD type. 50 x 50 mm (2" x 2") field of view.
- DC4: Dalsa, CCD type. 50 x 100 mm (2" x 4") field of view.
- DC22: Bioptics, CMOS type. 50 x 50 mm (2" x 2") field of view.
- DC24: Bioptics, CMOS type. 50 x 100 mm (2" x 4") field of view.
- DC44: Bioptics, CMOS type. 100 x 100 mm (4" x 4") field of view.
- DC3: Hamamatsu, CCD type. 24 x 34 mm (0.94" x 1.3") field of view.
- DC5: Hamamatsu C9730DK-11, CMOS type. 50 x 50 mm (2" x 2") field of view.
- DC10: Hamamatsu, CMOS type. 100 x 100 mm (4" x 4") field of view.
- DC12: Hamamatsu C9732DK, CMOS type. 120 x 120 mm (4.7" x 4.7") field of view.
Power supply
The original power supply is roughly:
- 12V in?
- Dual channel
- Power-one DGP12U5D5, WTT250IG
- Outputs +/- 5V
- Is the -5V rail used?
- Seem to output 5.3 or 5.4V
- LM317
- So basically: 12V DC is converted to just above 5V. Then a linear regulator smooths the last remaining bit down to 5V
Weidmüller-4 | Wire color | Volt | Note |
---|---|---|---|
1 | Blue | 5.11 | Ground? |
2 | Green | 0 | -5V? |
3 | Black | 0 | -5V? |
4 | Orange | 10.85 | +5V? |
DB9 | Wire | Weidmüller-8 | Volt |
---|---|---|---|
1 | Yellow | 2 | 0V |
2 | Blue | 2 | 0V |
3 | Gray | 4 | 0V |
4 | Black | 4 | 0V |
5 | Red | N/C | |
6 | Green | 1 | 5.15V |
7 | Violet | 1 | 5.15V |
8 | White | 3 | 5.05V |
9 | Brown | 3 | 5.05V |
N/C | 5 | ||
N/C | 6 | ||
N/C | 7 | ||
N/C | 8 |
Weidmüller BL series 3.5 mm connectors
MX-20
Cabinet x-ray
Camera
- DC-12: 12 cm x 12 cm
- DC-5: 5 cm x 5 cm, less expensive
10 lp/mm spatial resolution in contact mode. Magnification allows for spatial resolution of up to 50 lp/mm. The 20 μm nominal focal spot and up to 5X additional geometric magnification
Specs
- Window: 0.01” beryllium
- Resolution: Up to 50 lp/mm (10 lp/mm contact mode)
- 10-35 kVp @ 300 μA
- <20 μm spot size
- Image acquisition time: 15 sec
https://www.eevblog.com/forum/reviews/faxitron-mx-20-x-ray-system-teardown/
https://www.eevblog.com/forum/reviews/faxitron-mx-20-reverse-engineering-(with-some-good-results)/
https://www.youtube.com/watch?v=mYrjOIiJv88
VersaVision
Circa 2017?
Perkin-elmer_xrd-1640_an19-ets-lp
"The MX-20 had an Perkin Elmer panel sensor with parallel output to an SCSI PC interface board. It can do medium sized boards, up to 10" x 12"...The P-E sensor that came with my MX-20 was not it's original sensor. My MX-20 was non-digital, just a film based one (or phosphor plate based). The P-E had a low voltage differential, parallel, output that was sent through a cable to a PCI frame capture board. The board (made by Perkin Elmer) looked like a SCSI interface to the PC. The sensors are available on eBay, sometimes for a good price, but I've not seen a complete system, sensor, power supply, cables, interface card, and software, in a long time. A sensor is on eBay now, see"
tvrsales2 https://www.ebay.com/itm/PERKIN-ELMER-DIGITAL-X-RAY-DETECTOR-SYSTEM-XRD-1640/233202476578
PERKIN ELMER DIGITAL X-RAY DETECTOR SYSTEM XRD-1640
Teledyne dalsa CD26M112 X1
Dalsa Teledyne Xineos CMOS Xray Detector 130mm X 130mm CD26M112 X1
Misc
Jul 31, 2018 Hologic acquired Faxitron Bioptics for $85,000,000
Tweets
- Trying to recover CD data: https://twitter.com/johndmcmaster/status/1214059627576360960?s=20
- On ham power supply: https://twitter.com/johndmcmaster/status/1214058462591631360?s=20
- DC5 vs DC12 cover off: https://twitter.com/johndmcmaster/status/1214056617336926208?s=20
- DC internals: https://twitter.com/johndmcmaster/status/1214055732741435392?s=20