Difference between revisions of "Hamamatsu source"

From Nucwiki
Jump to: navigation, search
(Created page with "Head controllers are yellow Hamamatsu Photonics Softex L7452 X-Ray Tube Unit L7452 HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)")
 
Line 4: Line 4:
  
 
HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)
 
HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)
 +
 +
* Pony controller
 +
 +
"Pony Hamamatsu DR2 X-ray System X-RAYCONTROL Unit for PXG-090W & Cables"
 +
 +
Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on
 +
 +
 +
** serial
 +
 +
25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter
 +
 +
 +
** interlock1
 +
 +
2 wires. Came with shorting plug
 +
 +
Internally appear to go to J1
 +
 +
Why isn't interlock1 and interlock2 in series? Do they serve different purposes?
 +
 +
** interlock2
 +
 +
4 wires used in case
 +
* 1: red
 +
* 2: orange
 +
* 3: white
 +
* 4: gray
 +
 +
Red and orange go to J11 near large MATSUSHITA relay
 +
 +
Think white and gray go to J9 along with a few other wires
 +
 +
 +
** Internals
 +
 +
SW1: all LEDs on front light when pressed. No serial output
 +
 +
** MCU
 +
 +
MCU: TMP96C141BF
 +
 +
    $ strings c8033-11_v1.0.5.p9_2012-01-06.bin
 +
    FOXf
 +
    ??n1
 +
    Micro Foucus X-Ray 130kV Controller
 +
    Version 1.03p 00/03/25 sat.
 +
    KOU PLANNING  CO.LTD.
 +
    @h1!
 +
    ?9sp
 +
    ?6w^
 +
    ?9cI
 +
    ?9c8
 +
    ?9c'
 +
    HIV CUR PHV PCU AGN RST
 +
    ERR
 +
    ERROR
 +
    X-RAY ON
 +
    X-RAY OFF
 +
    HIV
 +
    CUR
 +
    PHV
 +
    PCU
 +
    OVER
 +
    NORMAL
 +
    INTER LOCK
 +
    AGING
 +
    AGING
 +
    READY
 +
    NOT READY
 +
    BUSY
 +
    REMOTE
 +
    LOCAL
 +
    RESPONSE
 +
    STANDBY
 +
    CAMERA POWER
 +
    SMALL
 +
    LARGE
 +
    Micro Foucus X-Ray Controller
 +
    Version 1.03p 00/03/25 sat.
 +
    KOU PLANNING  CO.LTD.
 +
    Control commands
 +
    HIV_0-1300    CUR_0-300
 +
    PHV_0-1300    PCU_0-300
 +
    AGN_N_N      XON    XOF
 +
    CXT    RST    CFL    CFS
 +
    Status commands
 +
    STX    SHV    SCU    SPV
 +
    SPC    SOV    SIN    SAG
 +
    STS    SRB    SRL    DAG
 +
    9;1d
 +
    DRSLVMFHGN!BTO$x
 +
    x= x
 +
    gH[;(
 +
    ?.fE
 +
    <8Db
 +
    @0000
 +
    address : +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +A +B +C +D +E +F  0123456789ABCDEF
 +
    ----------------------------------------------------------  ----------------
 +
      PCR      XWA      XBC      XDE      XHL      XIX      XIY      XIZ      XSP
 +
    Version 0.04a  94/06/02 thu.
 +
    ERROR
 +
    xwa:
 +
    xbc:
 +
    xde:
 +
    xhl:
 +
    xix:
 +
    xiy:
 +
    xiz:
 +
    xsp:
 +
    pcr:
 +
      sr:
 +
      spend:
 +
    usec
 +
    TLCS-900 Remote Debugger  Version 0.04a
 +
    KOU PLANNING CO. LTD.    94/06/02 thu.
 +
    D[(start),(end)]      :  memory dump
 +
    R                      :  register display
 +
    S[add]                :  set memory
 +
    G[(add),(break)]      :  jump to add.
 +
    N[(add),(break)]      :  jump to add.
 +
    L[offset add]          :  load file
 +
    M(source),(end),(dist) :  move memory
 +
    F(start),(end),(data)  :  fill memory
 +
    H                      :  help massage
 +
    V                      :  version
 +
    ?.fE

Revision as of 21:54, 1 June 2022

Head controllers are yellow

Hamamatsu Photonics Softex L7452 X-Ray Tube Unit L7452

HAMAMATSU PHOTONICS L7902 (head) & C8033-01 (controller)

  • Pony controller

"Pony Hamamatsu DR2 X-ray System X-RAYCONTROL Unit for PXG-090W & Cables"

Couldn't get x-ray on. While interlock1 came shorted, interlock2 is not. Suspect that is the issue and a yellow light will come on


    • serial

25 pin serial. Used a DB25 to DB9 adapter and was able to get the message "ERROR" when I pressed enter


    • interlock1

2 wires. Came with shorting plug

Internally appear to go to J1

Why isn't interlock1 and interlock2 in series? Do they serve different purposes?

    • interlock2

4 wires used in case

  • 1: red
  • 2: orange
  • 3: white
  • 4: gray

Red and orange go to J11 near large MATSUSHITA relay

Think white and gray go to J9 along with a few other wires


    • Internals

SW1: all LEDs on front light when pressed. No serial output

    • MCU

MCU: TMP96C141BF

   $ strings c8033-11_v1.0.5.p9_2012-01-06.bin 
   FOXf
   ??n1
   Micro Foucus X-Ray 130kV Controller
   Version 1.03p 00/03/25 sat.
   KOU PLANNING  CO.LTD.
   @h1!
   ?9sp
   ?6w^
   ?9cI
   ?9c8
   ?9c'
   HIV CUR PHV PCU AGN RST
   ERR 
   ERROR
   X-RAY ON
   X-RAY OFF
   HIV 
   CUR 
   PHV 
   PCU 
   OVER
   NORMAL
   INTER LOCK 
   AGING 
   AGING
   READY
   NOT READY
   BUSY
   REMOTE
   LOCAL
   RESPONSE 
   STANDBY
   CAMERA POWER 
   SMALL
   LARGE
   Micro Foucus X-Ray Controller
   Version 1.03p 00/03/25 sat. 
   KOU PLANNING  CO.LTD.
   Control commands
   HIV_0-1300    CUR_0-300
   PHV_0-1300    PCU_0-300
   AGN_N_N       XON    XOF
   CXT    RST    CFL    CFS
   Status commands
   STX    SHV    SCU    SPV
   SPC    SOV    SIN    SAG
   STS    SRB    SRL    DAG
   9;1d
   DRSLVMFHGN!BTO$x
   x=	x
   gH[;(
   ?.fE
   <8Db
   @0000
    address : +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +A +B +C +D +E +F  0123456789ABCDEF
   ----------------------------------------------------------  ----------------
      PCR      XWA      XBC      XDE      XHL      XIX      XIY      XIZ      XSP
   Version 0.04a  94/06/02 thu.
   ERROR
    xwa:
    xbc:
    xde:
    xhl:
    xix:
    xiy:
    xiz:
    xsp:
    pcr:
     sr:
      spend:
   usec
   TLCS-900 Remote Debugger  Version 0.04a
   KOU PLANNING CO. LTD.     94/06/02 thu.
   D[(start),(end)]       :  memory dump
   R                      :  register display
   S[add]                 :  set memory
   G[(add),(break)]       :  jump to add.
   N[(add),(break)]       :  jump to add.
   L[offset add]          :  load file
   M(source),(end),(dist) :  move memory
   F(start),(end),(data)  :  fill memory
   H                      :  help massage
   V                      :  version
   ?.fE